Bơm tiêm công nghiệp
IZS30-1A Industrial Syringe Pump OEM Microliquid Handling Manufacturer
SKU: IZS30-1A
Industrial syringe pump. Mainly used with devices or instruments, especially for programmable liquid handling pump modules. Solve high-precision dispensing pump for micro liquid.
#syringe pump
Chia sẻ
Control resolution
1 step or 0.01mm
Valve type
3 Port 120º Valve
Width
44.4mm
Depth
140mm
Product Description
Industrial syringe pump. Mainly used with devices or instruments, especially for programmable liquid handling pump modules. Solve high-precision dispensing pump for micro liquid.
Working principle
Stepper motor drive, ball screw drive, rotary encoder to monitor displacement, with 0-bit detection
Rated stroke
30mm corresponds to 3000 steps
Linear speed range
50μm/sec – 5mm/sec (ie: 20 minutes slower for full stroke; 1.2 seconds faster for full stroke)
Control resolution
1 step or 0.01mm
Stroke control accuracy
When the stroke is greater than or equal to 30%, the error is less than or equal to 5‰
Piston driving force
Maximum driving force ≥ 7kg; secondary driving force ≥ 3kg
Injector type
50μl ,100μl ,250μl, 500μl, 1ml ,2.5ml ,5ml
Valve type
3 Port 120º Valve
Transposition time
Transposition time of two adjacent ports≤250ms
Valve drive
Stepper motor drive, photoelectric encoder feedback position, out-of-step monitoring
Valve material
Spool: Teflon Body: Teflon
Valve interface
Tubing fitting: 1/4-28 thread fitting; Injector fitting: 1/4-28 thread fitting
Communication Interface
RS485 communication interface, communication rate: 9600 baud rate/38400 baud rate optional
External input interface
2 external TTL inputs with isolation; used to control start-stop after pause
External output interface
3 external TTL outputs with isolation
Device address dial interface
The device address can be set externally through the BCD dial
Parameter jumper selection interface
Reserve 6 external jumper selection parameters (valve selection, communication rate selection, communication mode selection)
Initialization command series
Complete the initialization state of valve and piston through various commands
Parameter setting command series
Complete the setting of parameters such as speed, start slope, relative 0 point position, dead zone volume, etc. through different commands
Valve Control Command Series
Complete movement of various valve positions
Piston Control Command Series
Accurate displacement of the piston by different commands
Control command series
Single, repeated, and delayed execution of single commands, combined command sequences, and stored command sequences through different commands; as well as suspending and terminating tasks, etc.
Report command series
Through various commands, the valve position, piston position, piston driving force, piston speed and other states of the equipment can be monitored in real time
Width
44.4mm
Depth
140mm
Current
≥1.5A
Humidity
20%-95% at 40℃
Parameter Table
| Piston | |
|---|---|
| Working principle | Stepper motor drive, ball screw drive, rotary encoder to monitor displacement, with 0-bit detection |
| Rated stroke | 30mm corresponds to 3000 steps |
| Linear speed range | 50μm/sec – 5mm/sec (ie: 20 minutes slower for full stroke; 1.2 seconds faster for full stroke) |
| Control resolution | 1 step or 0.01mm |
| Stroke control accuracy | When the stroke is greater than or equal to 30%, the error is less than or equal to 5‰ |
| Piston driving force | Maximum driving force ≥ 7kg; secondary driving force ≥ 3kg |
| Injector type | 50μl ,100μl ,250μl, 500μl, 1ml ,2.5ml ,5ml |
| Applicable valve | |
| Valve type | 3 Port 120º Valve |
| Transposition time | Transposition time of two adjacent ports≤250ms |
| Valve drive | Stepper motor drive, photoelectric encoder feedback position, out-of-step monitoring |
| Valve material | Spool: Teflon Body: Teflon |
| Valve interface | Tubing fitting: 1/4-28 thread fitting; Injector fitting: 1/4-28 thread fitting |
| External control interface | |
| Communication Interface | RS485 communication interface, communication rate: 9600 baud rate/38400 baud rate optional |
| RS232 communication interface, communication rate: 9600 baud rate/38400 baud rate optional | |
| External input interface | 2 external TTL inputs with isolation; used to control start-stop after pause |
| External output interface | 3 external TTL outputs with isolation |
| Device address dial interface | The device address can be set externally through the BCD dial |
| Parameter jumper selection interface | Reserve 6 external jumper selection parameters (valve selection, communication rate selection, communication mode selection) |
| Software function | |
| Initialization command series | Complete the initialization state of valve and piston through various commands |
| Parameter setting command series | Complete the setting of parameters such as speed, start slope, relative 0 point position, dead zone volume, etc. through different commands |
| Valve Control Command Series | Complete movement of various valve positions |
Parameter Table
| Industrial syringe pump IZS30-1A applicable hose | ||
|---|---|---|
| Article number | Describe | Length |
| 008T16-050-20 | 1.6mm (1/16") O.D. x 0.5mm (.02") I.D. | 20m |
| 008T16-050-200 | 1.6mm (1/16") O.D. x 0.5mm (.02") I.D. | 200m |
| 008T16-100-20 | 1.6mm (1/16") O.D. x 1.0mm (.039") I.D. | 20m |
| 008T16-100-200 | 1.6mm (1/16") O.D. x 1.0mm (.039") I.D. | 200m |
| 008T32-150-10 | 3.2mm (1/8") O.D. x 1.5mm (.059") I.D. | 10m |
| 008T32-150-100 | 3.2mm (1/8") O.D. x 1.5mm (.059") I.D. | 100m |